Pickling lines are commonly used to remove oxide scale, activate, and clean the surfaces of steel, aluminum, and components. Media may include hydrochloric acid, sulfuric acid, nitric acid, hydrofluoric acid, or blended acid solutions. Differential pressure sensors are responsible for monitoring filter differential pressure, tank liquid level, circulating pump inlet and outlet differential pressure, and pipeline resistance. They are important measuring components for stable control of the pickling process.
A so-called differential pressure sensor for strong acids is not simply a standard differential pressure transmitter with an external corrosion-resistant enclosure. It requires integrated adaptation of the isolation diaphragm, sealing structure, pressure tapping method, fill fluid, wetted materials, and installation location. Improper selection in any part may result in zero drift, delayed response, output fluctuations, diaphragm perforation, or signal interruption.
Frequent sensor failures on site are often not caused by a single product quality issue, but by the combined effects of acid concentration, temperature, impurities, pressure surges, and maintenance practices. Only by accurately identifying the root causes of failure can the risk of shutdown maintenance be reduced and pickling process parameters be kept within a controllable range.
The core wetted component of a differential pressure sensor is the isolation diaphragm. Ordinary 316L stainless steel can be used for a short period in certain low-concentration acidic media. However, when exposed to high-concentration hydrochloric acid, chloride-containing acid solutions, or hydrofluoric acid, it may develop pitting corrosion, crevice corrosion, and stress corrosion. As the diaphragm is generally thin, intensified corrosion can rapidly reduce the sensor's long-term stability.
For example, chloride ions in hydrochloric acid environments have strong penetration capability; sulfuric acid applications are closely related to concentration and temperature; and fluorine-containing media may pose particular risks to glass, ceramics, and certain metal materials. Therefore, selection cannot be based solely on a general description of “acid resistance”; the acid type, mass fraction, maximum temperature, and impurity composition must be clearly defined.
For highly corrosive applications, Hastelloy, tantalum, PTFE lining, PVDF, PFA, or remote diaphragm seal solutions should be considered comprehensively. Wetted materials must be selected based on actual media compatibility rather than simply pursuing a higher-grade stainless steel material.
Differential pressure sensor failure is not always caused by diaphragm damage. If O-rings, gaskets, pressure fittings, or cable seals are made of materials that are not resistant to the medium, swelling, hardening, or cracking may cause microleakage. Once acid mist enters the enclosure, it may corrode terminals, circuit boards, and solder joints, resulting in intermittent failures.
Pickling areas generally have high humidity and heavy acid mist. Even if a sensor enclosure is rated IP65, this does not mean it can withstand long-term exposure to acidic condensate. Cable entries, terminal box covers, and vent holes are vulnerable locations that are easily overlooked and should be fitted with compatible corrosion-resistant sealing accessories.
During on-site maintenance, failure to replace seals after disassembly and reassembly, failure to tighten threaded connections to the specified torque, or crystallized residues left on sealing surfaces can all render an otherwise reliable protective structure ineffective. In highly corrosive areas, it is recommended to arrange the electronics compartment separately from the media side to reduce direct exposure to acid mist.
Pickling solutions often contain iron oxide scale, metal debris, sediment, flocculants, and reaction products. When differential pressure measurement uses small pressure tapping holes, capillary tubes, or standard impulse lines, impurities can easily accumulate in low-flow areas. After one side of the passage becomes partially blocked, the sensor measures delayed local pressure rather than the actual differential pressure.
Taking filter differential pressure monitoring as an example, if the high-pressure-side tapping hole is covered by sediment, the displayed value may be too low and the system may fail to indicate filter element blockage in time. If the low-pressure side is blocked, an abnormally high differential pressure alarm may occur. Both situations can affect circulating pump operation and acid solution filtration performance.
After the acid solution is shut down for cooling or its concentration changes, certain salts may crystallize inside the pressure tapping structure. Particularly during long production interruptions or significant ambient temperature variations, crystals may repeatedly form and detach, causing output jumps, hysteresis, or failure of the zero point to return.
After abnormal readings are detected, some on-site personnel use hard metal tools to clear pressure tapping holes, which can easily scratch the isolation diaphragm or damage the corrosion-resistant layer. Once fine scratches appear on the isolation diaphragm surface, corrosive media preferentially attack these defective areas, significantly accelerating subsequent failure.
A more appropriate approach is to use compatible flushing fluid, low-pressure cleaning, and standardized disassembly and inspection according to process conditions. After cleaning, the zero point, span output, and static pressure balance on both sides should be checked. The presence of an output signal alone should not be used as the criterion for determining that the equipment has returned to normal.
For measuring points where differential pressure rises slowly over time and rapidly recurs after repeated cleaning, the installation structure itself needs to be investigated. Normally, priority should be given to adjusting the pressure tapping location, increasing the effective passage cross-section, or using a diaphragm-sealed differential pressure measurement structure.
The start-stop operation of circulating pumps, rapid opening and closing of switching valves, and insufficient pipeline venting in pickling lines may all generate water hammer and transient pressure surges. Even if the normal sensor range is only 10 kPa, 25 kPa, or 100 kPa, its static pressure rating, one-sided overload capacity, and instantaneous pressure limit must be verified separately.
Differential pressure elements are most vulnerable to asymmetric pressure on both sides. When the pressure on the high-pressure side suddenly rises while the low-pressure side responds with a delay, the instantaneous differential pressure actually acting on the sensor capsule may be far higher than the normal measured value. Frequent impacts can cause fatigue in the sensing element, resulting in zero drift, degraded linearity, or reduced repeatability.
In terms of control strategy, impacts can be reduced by optimizing pump soft-start time, reducing valve actuation speed, setting up bypass pressure relief, and avoiding dry-pipe operation. For applications with low differential pressure but high pipeline static pressure, particular attention should be paid to the one-sided overload rating of high-static-pressure differential pressure sensors during selection.
Pickling tank solution temperature may rise from ambient temperature to 60℃, 80℃, or even higher, and then gradually decrease after equipment shutdown. The thermal expansion coefficients of the diaphragm, housing, sealing rings, and fill fluid differ. Long-term hot and cold cycling can change seal preload and increase stress in internal structures.
Temperature can also alter medium viscosity, crystallization tendency, and the zero characteristics of electronic components. When a sensor is installed near a tank and is alternately exposed to acid mist, thermal radiation, and wash water, the actual ambient temperature may be significantly higher than the normal operating conditions indicated on the instrument nameplate.
Therefore, media temperature, ambient temperature, and short-term peak temperature should all be confirmed during selection; normal operating temperature alone is not sufficient. Where necessary, capillary remote transmission, heat-dissipation mounting plates, or relocation of the transmitter electronics to areas with less corrosion and lower temperatures may be used.
Reliable selection should at least provide the acid solution name, concentration range, maximum temperature, solid particle conditions, normal differential pressure, maximum differential pressure, pipeline static pressure, connection dimensions, and installation orientation. For production lines using mixed acids, waste acid recovery solution, or frequently adjusted formulations, all possible chemical components should also be listed.
The range should not be selected simply to equal the normal differential pressure. A reasonable margin is generally required for fluctuations, alarms, and short-term surges, while avoiding an excessively large range that would result in insufficient resolution for small differential pressure signals. The 4 mA to 20 mA output, RS485 communication, and local display functions should also be confirmed consistently with the existing control system.
Xi'an Shenghongchuang Instrumentation Co., Ltd. can configure differential pressure transmitters, diaphragm seals, and intelligent digital display control instruments based on specific measuring points such as pickling line filtration, circulation, and liquid level interlocks. Its production base, Shaanxi Qinkong Sensor Technology Co., Ltd., has capabilities in the R&D, manufacturing, and non-standard customization of sensors and transmitters, and can provide structural matching for complex media.
It is recommended to include differential pressure sensors in the pickling line inspection checklist and regularly inspect diaphragm appearance, flange seals, cable entries, acid mist corrosion traces, and the condition of pressure tapping passages. For critical filter differential pressure points, trend recording can be configured to identify blockages and measurement abnormalities in advance through the rate of differential pressure change.
During calibration, the zero point, full scale, return error, and output stability should be checked separately. If zero drift continues to increase, output frequently fluctuates abruptly, corrosion spots appear on the instrument surface, or leakage occurs at sealing locations, the instrument should be taken out of service for inspection promptly to prevent inaccurate sensor readings from causing incorrect process judgments.
Reliable operation of differential pressure sensors for strong acids depends on coordinated control of materials, structure, installation, and maintenance. Before project design or equipment modification, existing pickling process parameters and failure records should be compiled, and professional technical personnel should be consulted to confirm wetted materials, pressure limits, and installation solutions, thereby configuring truly suitable differential pressure measurement products for critical measuring points.