How to Assess the Linearity of a Pressure Transmitter: An Analysis of Range and Calibration Methods
Added to Favorites:125

How can you determine whether a linear pressure transmitter meets your measurement requirements? This article explains nonlinearity specifications, range selection, and calibration methods to support technical evaluation.

1. Understanding nonlinearity specifications

Nonlinearity describes the degree of deviation from a straight line

A pressure transmitter converts a pressure signal into a current, voltage, or digital signal. Ideally, changes in pressure and output are linearly related. In practice, the output curve of a product may be slightly curved, and the nonlinearity specification describes the degree of this deviation.

A common specification is “±0.1%FS,” where FS usually refers to the full-scale output span. For a 4–20mA output, for example, the output span is 16mA, so ±0.1%FS corresponds to an output deviation of ±0.016mA. For the specific determination, refer to the calculation method and test conditions stated in the product documentation.

Even when two products are both specified as ±0.1%FS, different manufacturers may use the endpoint straight-line method or the best-fit straight-line method. When comparing products, first confirm the reference line, test points, increasing and decreasing pressure procedures, and temperature conditions. Otherwise, the same specification may not represent equivalent test results.

Nonlinearity and total accuracy are not the same

Nonlinearity reflects only the deviation of the output curve from a reference line and does not represent the instrument’s total error on its own. Hysteresis, repeatability, zero offset, and temperature changes may also affect actual measurement results.

Some product documentation combines linearity, hysteresis, and repeatability under a total accuracy specification. When purchasing, confirm which error terms are included and whether temperature drift and long-term stability are covered. This helps avoid mistaking total accuracy for the nonlinearity specification alone.

For equipment that must operate reliably over the long term, consider not only the as-delivered specifications but also the operating temperature, on-site vibration, power supply quality, and scheduled calibration arrangements. Evaluate measurement performance against actual operating conditions.

2. How range selection affects effective measurement error

Consider both normal and maximum pressure

When selecting a model, first identify the type of pressure to be measured, such as gauge, absolute, or differential pressure. Then confirm the normal operating range, possible peak pressure, and process media conditions. The range must cover normal measurements as well as pressure changes during processes such as startup, shutdown, and valve operation.

With a range that is too wide, the allowable error calculated as a percentage of full scale may account for a significantly larger proportion of the reading at low pressure. A range that is too narrow may not cover pressure peaks and could expose the sensor to overpressure. The allowable overpressure is a specified limit the product can withstand; it does not mean the product can measure accurately at that pressure over the long term.

For example, two transmitters are both specified at ±0.1%FS, with ranges of 0–1MPa and 0–10MPa, respectively. Assuming all other conditions are the same and using this specification for the calculation, the results are as follows:

RangePressure corresponding to ±0.1% FSReference proportion at the 0.5 MPa measurement point
0~1MPa±0.001MPaApproximately ±0.2% of reading
0~10MPa±0.01MPaApproximately ±2% of reading

These values are only intended to illustrate the difference between a percentage of full scale and a percentage of the reading. They do not represent the total error of the equipment in the field. If approximately 0.5MPa is a critical measurement range, prioritize evaluating a better-matched range while ensuring it can accommodate peak pressure.

For low differential pressure applications, also check the process media

In applications such as duct pressure measurement and equipment ventilation monitoring, the differential pressure signal may be only a few hundred pascals. Leaks, water accumulation, or blockages in the pressure tubing can all affect the reading. When selecting a model, check not only the range but also whether the gas is clean and dry, and whether the pressure port connections are suitable for the site.

For example, the AZ-808 domestic differential pressure transmitter, wind pressure sensor, and gas micro differential pressure transmitter can be used to measure low differential pressures in non-corrosive, dust-free, dry gases. It offers 4–20mA and various voltage output options; confirm the specific configuration against the order specifications.

The product documentation for this series lists ranges from 0–500Pa to 700kPa, as well as total accuracy options of ±0.1%FS, ±0.25%FS, and ±0.4%FS. This does not mean that every range is available with all accuracy ratings or output configurations. Before selecting a model, confirm the available combinations and corresponding test conditions with the supplier.

3. How to check linearity during calibration

Prepare a reference instrument and a stable pressure source

Before calibration, confirm that the range and measurement capability of the standard pressure equipment are suitable for the transmitter under test, and assess measurement uncertainty in light of the required results. For low differential pressure calibration in particular, check fitting seals, the condition of the pressure tubing, and the effects of ambient airflow.

Connect the power supply and output measurement equipment as specified for the product, and confirm that the power supply, current loop, or voltage input is operating correctly. Allow the instrument to warm up, and ensure that its mounting orientation and ambient temperature comply with the manufacturer’s instructions. Record the actual conditions if necessary.

Before taking formal readings, perform pre-pressurization and checks in accordance with the applicable calibration procedure or product instructions. Increase pressure steadily and wait for the pressure and output to stabilize at each point before taking a reading. This avoids mistaking dynamic fluctuations for static nonlinearity error.

Increase and decrease pressure at multiple points, and record the results

A common practice is to set test points near 0%, 25%, 50%, 75%, and 100% of the range, first increasing the pressure point by point and then decreasing it point by point. To examine the curve in more detail, add test points, especially in the pressure range commonly used in practice.

For a 0–1MPa, 4–20mA transmitter, the theoretical output can be checked against the table below. For actual evaluation, use the actual pressure measured by the reference instrument to calculate the corresponding theoretical value, rather than relying only on the target pressure.

Span ratioPressureTheoretical Output
0%0MPa4mA
25%0.25MPa8mA
50%0.5MPa12mA
75%0.75MPa16mA
100%1MPa20mA

Record the actual output and deviation at each point, and compare the increasing- and decreasing-pressure data separately. The maximum indication error is not necessarily the nonlinearity error. Calculate nonlinearity using the agreed reference line and method, and determine hysteresis by comparing the increasing- and decreasing-pressure results at the same pressure point.

4. Set acceptance requirements based on site conditions

Consider temperature drift, stability, and installation together

Good linearity under laboratory conditions does not mean that field error will remain unchanged. When the temperature differs from the calibration conditions, the zero point and sensitivity may change. For applications with a wide temperature range, check the product’s compensated temperature range and the relevant drift specifications.

For example, the AZ-808 documentation lists a compensated temperature range of -10℃ to 55℃ and a typical zero-point temperature drift of 0.02%FS/℃. A typical value is not the guaranteed upper limit for every product. For an actual evaluation, also check the maximum value, applicable conditions, and total system error requirements.

During installation, check the positive and negative pressure ports, wiring polarity, and supply voltage range, and prevent the pressure tubing from being pressurized, accumulating liquid, or leaking. For hazardous areas requiring explosion protection, verify the validity of the certificate for the purchased configuration, the applicable area, and the requirements for associated safety barriers. Do not rely solely on product promotional text.

Create a practical purchasing and recalibration checklist

For technical discussions, compile the pressure type, normal range, peak pressure, process media, temperature, output signal, and allowable error. Specify whether accuracy is calculated as a percentage of full scale or of the reading. Also ask the supplier to explain the nonlinearity test method and the items included in the total accuracy specification.

At acceptance, retain the product documentation and calibration records, and confirm that the actual model and range match the order requirements. Set the recalibration interval based on process risk, operating environment, historical drift, and applicable metrology management requirements. Inspect the equipment promptly if an abnormal deviation or overpressure occurs.

To assess a specific application, provide Xi'an Shenghongchuang Instrumentation Co., Ltd. with the pressure range, process media conditions, mounting interface, and output requirements. Its technical staff can assist in checking the range, accuracy, and calibration plan.