Pressure Transmitter Manufacturer
Consultation hotline:15529283736
Product Center
—— Product Center ——
Xi'an Shenghongchuang Instrument Co., Ltd.
Contact: Mr. Zhang
Mobile: 15529283736
Email: shc-sensor@qq.com
Address: Fortune Building, Sanqiao Street, Xixian New Area, Xi'an, Shaanxi Province
Xi'an Shenghongchuang Instrumentation Co., Ltd.
Contact: Mr. Zhang
Mobile: 15529283736
Email: shc-sensor@qq.com
The capacitive intelligent pressure/differential pressure/level transmitter is a high-precision intelligent instrument product independently researched, developed, and manufactured by our company based on many years of scientific research and practical experience.
The intelligent pressure/differential pressure/level transmitter is a new generation of intelligent transmitters manufactured using a sensor core produced with ROSEMOUNT technology, combined with advanced digital technology developed by our company. In principle, it uses a direct digital capacitance circuit to replace the analog signal amplification circuit and A/D conversion circuit, and applies digital compensation technology to compensate for temperature and static pressure, thereby improving measurement accuracy and reducing temperature drift. The complete transmitter features small size, high reliability, and good long-term stability. This product offers an extremely high cost-performance ratio and has therefore become a mainstream product in the transmitter market.
This intelligent pressure/differential pressure/level transmitter features a novel design principle, complete varieties and specifications, easy installation and operation, and safety explosion-proof performance. It is especially well known for high accuracy, small size, light weight, convenient adjustment, good long-term stability, strong unidirectional overload protection capability, and suitability for use in harsh environments. Therefore, it enjoys a high reputation in domestic and international markets. This 0.1-class capacitive intelligent transmitter series includes pressure, differential pressure, absolute pressure, remote transmission, level, and other types, and is equipped with corresponding optional products. It can be widely used in electric power, metallurgy, petrochemical, instrumentation processing, and other industries, and complies with the "IEC" standard requirements promoted in China.
Product Features
·The use of a microprocessor provides greater flexibility and enhanced functions;
·Strong self-diagnostic capability;
·Wide measuring range coverage 0~0.2kPa…10MPa;
·Accuracy up to 0.1 class, rangeability better than 40:1;
·Zero point and span adjustments do not affect each other;
·Supports both remote and local adjustment of range and zero point;
·Two-wire system, compliant with HART protocol, capable of digital communication with ROSEMOUNT275 and MS-H376 handheld communicators without interrupting analog output;
·The sensor has built-in non-volatile memory;
·Good stability, high accuracy, adjustable damping, and strong resistance to unidirectional overload;
·No mechanical transmission parts, low maintenance workload, sturdy and vibration-resistant;
·All universal parts, interchangeability between the sensor and electronic circuit board does not affect transmitter performance, making maintenance more convenient;
·Optional diaphragm materials for parts in contact with the measured medium, explosion-proof housing structure;
·Industry-proven superiority and reliability.
Performance Parameters
Technical Performance
Applicable media: liquid, gas, or steam
Measuring range: Range 3 0-0.2~6.0kPa
Range 4 0-6.0~40.0kPa
Range 5 0-40.0~200kPa
Range 6 0-0.16~1.0MPa
Range 7 0-0.4~2.0MPa
Range 8 0-1.6~10MPa
Output signal: linear output: 4~20mA output superimposed with HART protocol digital signal (two-wire system);
Square root output: when the input pressure is between 1.0%~100%, the output has a square root relationship with differential pressure; between 0%~0.1%, the output has a segmented linear relationship with differential pressure; there is no abrupt change in the output between 4~20mADC, superimposed with HART? protocol digital signal (two-wire system), and users can select linear output or square root output according to on-site conditions.
Power supply: externally supplied 24V DC, (power supply range 14.5V~45V)
Load characteristics: the maximum allowable load resistance of the load resistance loop is Rmax=(E-10.5)/0.0236(Ω)
Installation in hazardous locations: flameproof Exd Ⅱ CT6;
intrinsically safe Exia Ⅱ CT6;
Load characteristics: after positive and negative migration, the upper and lower range limits shall not exceed the range limits. At the minimum range, the maximum positive migration is 0.975URL; the maximum negative migration is -URL. Under square root output, positive and negative migration can be adjusted and calibrated to 10% of the flow range.
Temperature range: medium temperature -40~100℃
storage temperature -40~100℃
operating temperature of the complete unit -40~85℃
Relative humidity: 0~100%
Overpressure limit: applying 0(absolute pressure)~13.76MPa pressure to either side of the transmitter will not damage the transmitter; the flange can withstand 68.9MPa pressure; normal operating pressure should be between 3.45kPa(absolute pressure) and the upper limit of the transmitter range.
Volumetric change: less than 0.16cm3
Damping: the time constant is adjustable between 0.2~32.0s.
Start-up time: 3s, no preheating required.
Performance indicators:
(Under standard test conditions with no migration, 316 stainless steel isolation diaphragm, and other standard conditions.)
Rangeability: 40:1
Accuracy: ranges 3, 4, 5
when the range ratio is 1:1 to 10:1, it is ±0.1% of the calibrated range,
when the range ratio is 10:1 to 40:1, it is ±0.05(1+0.1URL/range)% range,
ranges 6, 7, 8
when the range ratio is 1:1 to 10:1, it is ±0.15% of the calibrated range,
when the range ratio is 10:1 to 40:1, it is ±0.075(1+0.1URL/range)% range,
Stability: the error over six months is ±0.15% of the maximum range;
Temperature effect: for ranges 4, 5, 6, 7, 8
zero error is ±0.25% of the maximum range/55℃.
the total error including span and zero point is ±0.5% of the maximum range/55℃.
Note: for range 3, the temperature error should be doubled.
Overpressure effect: after applying pressure of 140kgf/cm2, the error is ±0.25% of the maximum range.
Static pressure effect: under linear output, zero error: after applying static pressure of 140kgf/cm2, the zero error for ranges 4 and 5 is ±0.25% of the maximum range, and the zero error for ranges 3, 6, 7, and 8 is ±0.5% of the maximum range.
Span error: the error for ranges 4, 5, 6, 7, and 8 is -1±0.25%/6.89MPa of the maximum range, and the error for range 3 is -1.5±0.25%. This is a systematic error and can be eliminated by calibrating according to the actual static pressure before installation.
Power supply effect: less than 0.005%/V of output range
Vibration effect: on any axis, at a frequency of 200Hz, the resulting error is ±0.05%/g of the maximum range.
Load effect: as long as the input voltage to the transmitter is higher than 14.5V, there is no load effect within the load operating area.
Installation position effect: it may produce a zero shift error not greater than 0.25 kPa, but it can be corrected and has no effect on the range;
The rotation of the measuring body relative to the flange has no effect.
Electromagnetic radiation: complies with IEC801 standard
Appearance and Structure
Structural materials:
Isolation diaphragm and vent/drain valve: 316 stainless steel, Hastelloy C, Monel;
Flange and joint: 316 stainless steel, Hastelloy C, Monel;
“O” ring (in contact with measured medium): fluororubber, nitrile rubber, ethylene-propylene;
Fill fluid: silicone oil;
Bolts: carbon steel galvanized, 1Cr18Ni9, 0Cr17Ni4CuNb, 42CrMo;
Electrical housing: low-copper aluminum alloy;
Coating: polyurethane epoxy resin;
Pressure connection fitting: for ranges 3, 4, and 5, the center hole distance of the threaded connection of the two flanges is 1/4-18NPT and 54mm;
Joint 1/2-14NPT or M20X1.5 male thread; when using a ball cone sealing joint, the center distance is 50.8, 54, 57.2mm;
Weight: 3.5kg(excluding options)
Since its establishment, Xi'an Shenghongchuang Instrument Co., Ltd. has mainly been engaged in the R&D and production of pressure, level, temperature, differential pressure, digital display, weighing, and flow instruments. Main products include Swiss metallux ceramic piezoresistive sensors, self-developed ceramic capacitive pressure sensors, diffused silicon, ceramic piezoresistive, strain gauge, metal capacitive, ceramic capacitive pressure transmitters, remote pressure transmitters, high-temperature pressure transmitters, flush diaphragm pressure transmitters, anti-corrosion pressure transmitters, integrated/split level transmitters, anti-corrosion level transmitters, lightning-protection level transmitters, high-temperature guided-pressure level transmitters, straight-rod level transmitters, flange level transmitters, wind pressure transmitters, metal capacitive differential pressure transmitters, diffused silicon differential pressure transmitters, pressure controllers, thermocouple/RTD temperature transmitters, and ultrasonic level gauges.
Notes 1.All supplied products are provided with a product certificate and instruction manual, including product number, technical parameters, wiring, factory date, etc. Please check carefully to avoid misuse.2.Wiring should be carried out strictly in accordance with our company's instructions for use.3.This product is a precision instrument. Random disassembly, collision, dropping, or poking the sensor with sharp tools is prohibited.4.The transmitter can work immediately after being powered on, but the output becomes stable after preheating for 30 minutes.5.If any abnormality is found during use, turn off the power supply, stop using it, inspect it, or directly contact our company's technical department.6.During transportation and storage, restore the original packaging and store it in a cool, dry, and ventilated warehouse.7.The quality warranty period is 12 months (damage caused by improper use such as pressure overload or wiring errors is not covered by the quality warranty).8.Special products are customized products and do not participate in the 7-day no-reason return policy. Customized products have a delivery period and are non-returnable and non-exchangeable.