Domestic capacitive pressure transmitter NS-3051GP absolute pressure sensor differential pressure transmitter
Updated time:Feb 04, 2024
  • Domestic capacitive pressure transmitter NS-3051GP absolute pressure sensor differential pressure transmitter
  • Domestic capacitive pressure transmitter NS-3051GP absolute pressure sensor differential pressure transmitter
  • Domestic capacitive pressure transmitter NS-3051GP absolute pressure sensor differential pressure transmitter

Xi'an Shenghongchuang Instrumentation Co., Ltd.

Contact: Mr. Zhang

Mobile: 15529283736
Email: shc-sensor@qq.com

Contact Us

The capacitive intelligent pressure/differential pressure/level transmitter is a high-precision intelligent instrument product independently developed by our company based on many years of scientific research and practical experience.

The intelligent pressure/differential pressure/level transmitter is a new-generation intelligent transmitter manufactured using a sensor core produced with ROSEMOUNT technology, combined with advanced digital technology developed by our company. In principle, it uses a direct digital capacitance circuit to replace the analog signal amplification circuit and A/D conversion circuit, and adopts digital compensation technology to compensate for temperature static pressure, thereby improving measurement accuracy and reducing temperature drift. The complete transmitter features small size, high reliability, and good long-term stability. This product offers extremely high cost performance and has thus become a mainstream product in the transmitter market.

This intelligent pressure/differential pressure/level transmitter features a novel design principle, complete varieties and specifications, convenient installation and use, and safety explosion-proof characteristics. It is especially well known for high accuracy, small size, light weight, convenient adjustment, good long-term stability, strong one-way overload protection capability, and suitability for use in harsh environments. Therefore, it enjoys a very high reputation in domestic and international markets. This series of 0.1 grade capacitive intelligent transmitters includes multiple types such as pressure, differential pressure, absolute pressure, remote transmission, and level, and is equipped with corresponding optional products. It can be widely used in industries such as electric power, metallurgy, petrochemicals, and instrument processing, and complies with the "IEC" standard requirements implemented in China.


                                                                     


Product Features


·The use of a microprocessor provides greater flexibility and enhanced functions;

·Strong self-diagnostic capability;

·Wide range coverage 0~0.2kPa…10MPa;

·Accuracy up to 0.1 grade, turndown ratio better than 40:1;

·Zero point and span adjustment do not affect each other;

·Supports both remote and local adjustment of range and zero point;

·Two-wire system, compliant with HART protocol, capable of digital communication with ROSEMOUNT275 and MS-H376 handheld communicators without interrupting analog output;

·Built-in non-volatile memory in the sensor;

·Good stability, high accuracy, adjustable damping, and strong resistance to one-way overload;

·No mechanical transmission parts, low maintenance workload, sturdy and vibration-resistant;

·All universal parts, interchangeability between sensor and electronic circuit board does not affect transmitter performance, making maintenance more convenient;

·Optional diaphragm materials for parts in contact with the medium, explosion-proof housing structure;

·Industry-proven superiority and reliability.


Performance Parameters


Technical Performance

Applicable media: liquid, gas, or steam

Measuring range: Range 3  0-0.2~6.0kPa

Range 4  0-6.0~40.0kPa

Range 5  0-40.0~200kPa

Range 6  0-0.16~1.0MPa

Range 7  0-0.4~2.0MPa

Range 8  0-1.6~10MPa

Output signal: linear output: 4~20mA output superimposed with HART protocol digital signal(two-wire system);

Square root output: when the input pressure is between 1.0%~100%, the output has a square root relationship with differential pressure; between 0%~0.1%, the output has a segmented linear relationship with differential pressure; there is no jump change between 4~20mADC output, superimposed with HART? protocol digital signal(two-wire system), and users may select linear output or square root output according to on-site conditions.

Power      supply: externally supplied 24V DC,(power supply range 14.5V~45V)

Load characteristics: the maximum allowable load resistance in the load resistance circuit Rmax=(E-10.5)/0.0236(Ω)


                 


Installation in hazardous locations: flameproof Exd Ⅱ CT6;

               intrinsically safe Exia Ⅱ CT6;

Load characteristics: after positive or negative migration, the upper and lower limits of the range must not exceed the range limits. At the minimum range, the maximum positive migration is 0.975URL; the maximum negative migration is -URL. During square root output, positive and negative migration can calibrate 10% of the flow range.

Temperature range: medium temperature    -40~100℃

         storage temperature    -40~100℃

         complete unit operating temperature  -40~85℃

Relative humidity: 0~100%

Overpressure limit: applying 0(absolute pressure)~13.76MPa pressure to either side of the transmitter will not damage the transmitter; the flange can withstand 68.9MPa pressure; normal operating pressure should be between 3.45kPa(absolute pressure) and the upper limit of the transmitter range.

Volume change: less than 0.16cm3

Damping: adjustable time constant between 0.2~32.0s.

Startup time: 3s, no preheating required.


Performance indicators:

   (Under standard test conditions with no migration, 316 stainless steel isolating diaphragm, and other standard conditions.)

    Turndown ratio: 40:1

    Accuracy: ranges 3, 4, 5

                when the turndown ratio is 1:1 to 10:1, it is ±0.1% of calibrated span,

                      when the turndown ratio is 10:1 to 40:1, it is ±0.05(1+0.1URL/span)% span,

                ranges 6, 7, 8

                      when the turndown ratio is 1:1 to 10:1, it is ±0.15% of calibrated span,

                      when the turndown ratio is 10:1 to 40:1, it is ±0.075(1+0.1URL/span)% span,

    Stability: six-month error is ±0.15% of the maximum span;

    Temperature effect: ranges 4, 5, 6, 7, 8

                   zero error is ±0.25%/55℃ of the maximum span.

         the total error including span and zero point is ±0.5%/55℃ of the maximum span.

    Note: for range 3, the temperature error shall be doubled.

    Overpressure effect: after applying pressure of 140kgf/cm2, the error is ±0.25% of the maximum span range.

    Static pressure effect: during linear output, zero error: after applying static pressure of 140kgf/cm2, the zero error of ranges 4 and 5 is ±0.25% of the maximum span range, and the zero error of ranges 3, 6, 7, and 8 is ±0.5% of the maximum span range.

  Span error: the error of ranges 4, 5, 6, 7, and 8 is -1±0.25%/6.89MPa of the maximum span range, and the error of range 3 is -1.5±0.25%. This is a system error and may be eliminated by calibrating according to the actual static pressure before installation.

    Power supply effect: less than 0.005%/V of output span

    Vibration effect: on any axis, at a frequency of 200Hz, the induced error is ±0.05%/g of the maximum span range.

    Load effect: as long as the voltage input to the transmitter is higher than 14.5V, there is no load effect within the load operating area.

    Installation position effect: a zero position error not greater than 0.25 kPa may be generated at most, but it can be corrected, with no effect on span;

Rotation of the measuring body relative to the flange has no effect.

    Electromagnetic radiation: complies with IEC801 standard


Appearance Structure

 Structural materials:

        Isolation diaphragm and vent/drain valve: 316 stainless steel, Hastelloy C, Monel;

    Flange and connector: 316 stainless steel, Hastelloy C, Monel;

       “O” ring(in contact with measuring medium): fluororubber, nitrile rubber, ethylene-propylene;

    Filling liquid: silicone oil;

    Bolt: carbon steel zinc plated, 1Cr18Ni9, 0Cr17Ni4CuNb, 42CrMo;

        Electrical housing: low-copper aluminum alloy;

        Coating: polyester epoxy resin;

        Pressure connection fitting: for ranges 3, 4, and 5, the center hole distance of the threaded connection between two flanges is 54mm for 1/4-18NPT;

         Connector 1/2-14NPT or M20X1.5 male thread; when using conical seat sealing with connector, the center distance is 50.8, 54, 57.2mm;

        Weight:   3.5kg(excluding options)


                                                                                                       


Since its establishment, Xi'an Shenghongchuang Instrument Co., Ltd. has mainly engaged in the R&D and production of pressure, level, temperature, differential pressure, digital display, weighing, and flow instruments. Main products include Swiss metallux ceramic piezoresistive sensors, self-developed ceramic capacitive pressure sensors, diffused silicon, ceramic piezoresistive, strain gauge, metal capacitive, ceramic capacitive pressure transmitters, remote pressure transmitters, high-temperature pressure transmitters, flush diaphragm pressure transmitters, anti-corrosion pressure transmitters, integrated/split level transmitters, anti-corrosion level transmitters, lightning protection level transmitters, high-temperature guided pressure level transmitters, rigid rod level transmitters, flange level transmitters, air pressure transmitters, metal capacitive differential pressure transmitters, diffused silicon differential pressure transmitters, pressure controllers, thermocouple/RTD temperature transmitters, and ultrasonic level meters.



Notes 1.All supplied products are accompanied by product certificates and instructions, including product number, technical parameters, wiring, factory date, etc. Please carefully check them to avoid misuse. 2.Wiring should be carried out strictly in accordance with our company's instructions for use. 3.This product is a precision instrument. Random disassembly, collision, dropping, or probing the sensor with sharp tools is prohibited. 4.The transmitter can work immediately after power-on, but the output becomes stable after preheating for 30 minutes. 5.If any abnormality is found during use, the power supply should be turned off, use should be stopped, and inspection should be carried out, or directly contact our company's technical department. 6.During transportation and storage, restore the original packaging and store it in a cool, dry, and ventilated warehouse. 7.The quality warranty period is 12 months(damage caused by improper use such as pressure overload and incorrect wiring is not covered by the warranty). 8.Special products are customized products, do not participate in the 7-day no-reason return policy, and customized products have a delivery lead time, with no return or exchange.

u=2776578472,2395865860&fm=199&app=68&f=JPEG.jpg