Domestic Capacitance Pressure Transmitter HC3051DR High-Precision Intelligent Differential Pressure Transmitter
Updated time:Feb 18, 2024
  • Domestic Capacitance Pressure Transmitter HC3051DR High-Precision Intelligent Differential Pressure Transmitter
  • Domestic Capacitance Pressure Transmitter HC3051DR High-Precision Intelligent Differential Pressure Transmitter
  • Domestic Capacitance Pressure Transmitter HC3051DR High-Precision Intelligent Differential Pressure Transmitter

Xi'an Shenghongchuang Instrumentation Co., Ltd.

Contact: Mr. Zhang

Mobile: 15529283736
Email: shc-sensor@qq.com

Contact Us

The capacitive intelligent pressure/differential pressure/level transmitter is a high-precision intelligent instrument product independently researched, developed, and manufactured by our company on the basis of many years of scientific research and practical experience.

The intelligent pressure/differential pressure/level transmitter is a new-generation intelligent transmitter manufactured using a sensor core produced with ROSEMOUNT technology and integrated with the advanced digital technology developed by our company. In principle, it uses a direct digital capacitance circuit to replace the analog signal amplification circuit and A/D conversion circuit, and applies digital compensation technology to compensate for temperature and static pressure, thereby improving measurement accuracy and reducing temperature drift. The complete transmitter features small size, high reliability, and good long-term stability. With an extremely high cost-performance ratio, this product has become a mainstream product in the transmitter market.

This intelligent pressure/differential pressure/level transmitter features a novel design principle, complete varieties and specifications, easy installation and operation, safety, and explosion protection. It is especially known for high accuracy, compact size, light weight, convenient adjustment, good long-term stability, strong one-way overload protection capability, and suitability for use in harsh environments. Therefore, it enjoys a high reputation in both domestic and international markets. This series of 0.1 class capacitive intelligent transmitters includes pressure, differential pressure, absolute pressure, remote transmission, level, and other types, and is equipped with corresponding optional products. It can be widely used in industries such as electric power, metallurgy, petrochemical, and instrument processing, and complies with the IEC standards promoted in China.


                                                                     


Product Features


·The use of a microprocessor provides greater flexibility and enhanced functions;

·Strong self-diagnostic capability;

·Wide measuring range coverage 0~0.2kPa…10MPa;

·Accuracy up to 0.1 class, with a turndown ratio better than 40:1;

·Zero point and span adjustment do not affect each other;

·Supports both remote and local adjustment of span and zero point;

·Two-wire system, compliant with HART protocol, capable of digital communication with ROSEMOUNT275 and MS-H376 handheld communicators without interrupting analog output;

·The sensor is equipped with built-in non-volatile memory;

·Good stability, high accuracy, adjustable damping, and strong one-way overload resistance;

·No mechanical transmission parts, low maintenance workload, sturdy and vibration-resistant;

·All parts are universal, and interchange of the sensor and electronic circuit board does not affect transmitter performance, making maintenance more convenient;

·Optional diaphragm materials for contact with the measured medium, explosion-proof housing structure;

·Superiority and reliability proven by the industry.


Performance Parameters


Technical Performance

Applicable media:liquid, gas, or steam

Measuring range:Range 3  0-0.2~6.0kPa

Range 4  0-6.0~40.0kPa

Range 5  0-40.0~200kPa

Range 6  0-0.16~1.0MPa

Range 7  0-0.4~2.0MPa

Range 8  0-1.6~10MPa

Output signal:Linear output:4~20mA output superimposed with HART protocol digital signal(two-wire system);

Square root output:When the input pressure is between 1.0%~100%, the output has a square root relationship with differential pressure;between 0%~0.1%, the output has a segmented linear relationship with differential pressure;there is no jump change between 4~20mADC output, superimposed with HART? protocol digital signal(two-wire system), and users can choose linear output or square root output according to on-site conditions.

Power      supply:External power supply 24V DC, (power supply range 14.5V~45V)

Load characteristics:Maximum allowable load resistance in the load resistor loop Rmax=(E-10.5)/0.0236(Ω)


                 


Installation in hazardous locations: Flameproof type Exd Ⅱ CT6;

               Intrinsically safe type Exia Ⅱ CT6;

Load characteristics:After positive and negative migration, the upper and lower limits of the range must not exceed the limits of the range. At the minimum range, the maximum positive migration is 0.975URL;the maximum negative migration is -URL. During square root output, positive and negative migration can be adjusted to calibrate 10% of the flow range.

Temperature range:Medium temperature    -40~100℃

         Storage temperature    -40~100℃

         Complete unit operating temperature  -40~85℃

Relative humidity:0~100%

Overpressure limit:Applying 0(absolute pressure)~13.76MPa pressure to either side of the transmitter will not damage the transmitter;the flange can withstand 68.9MPa pressure;normal operating pressure should be between 3.45kPa(absolute pressure)and the upper range limit of the transmitter.

Volume change:less than 0.16cm3

Damping: Adjustable time constant between 0.2~32.0s.

Startup time:3s, no preheating required.


Performance indicators:

   (Under standard test conditions without migration, with 316 stainless steel isolation diaphragm and other standard test conditions.)

    Turndown ratio:40:1

    Accuracy:Range 3, 4, 5

                When the turndown ratio is 1:1 to 10:1, it is ±0.1% of the calibrated range,

                      when the turndown ratio is 10:1 to 40:1, it is ±0.05(1+0.1URL/range)% range,

               Range 6, 7, 8

                      When the turndown ratio is 1:1 to 10:1, it is ±0.15% of the calibrated range,

                      when the turndown ratio is 10:1 to 40:1, it is ±0.075(1+0.1URL/range)% range,

    Stability:maximum error over six months is ±0.15% of the maximum range;

    Temperature effect:Range 4, 5, 6, 7, 8

                   Zero point error is ±0.25%/55℃ of the maximum range.

         Total error including range and zero point is ±0.5%/55℃ of the maximum range.

    Note:For Range 3, the temperature error should be doubled.

    Overpressure effect:After applying pressure of 140kgf/cm2, the error is ±0.25% of the maximum range.

    Static pressure effect:During linear output, zero point error:after applying static pressure of 140kgf/cm2, the zero point error of Range 4 and 5 is ±0.25% of the maximum range, and the zero point error of Range 3, 6, 7, and 8 is ±0.5% of the maximum range.

  Range error:The error of Range 4, 5, 6, 7, and 8 is -1±0.25%/6.89MPa of the maximum range, and the error of Range 3 is -1.5±0.25%. This is a system error and can be eliminated by calibrating according to the actual static pressure before installation.

    Power supply effect:less than 0.005%/V of output range

    Vibration effect:On any axis, at a frequency of 200Hz, the induced error is ±0.05%/g of the maximum range.

    Load effect:As long as the input voltage to the transmitter is higher than 14.5V, there is no load effect within the load operating area.

    Mounting position effect:A zero shift error not greater than 0.25 kPa may be generated, but it can be corrected and has no effect on the range;

Rotation of the measuring body relative to the flange has no effect.

    Electromagnetic radiation:complies with IEC801 standard


External Structure

 Structural materials:

        Isolation diaphragm and vent/drain valve:316 stainless steel, Hastelloy C, Monel;

    Flange and connector:316 stainless steel, Hastelloy C, Monel;

        “O” ring(in contact with the measured medium):fluororubber, nitrile rubber, ethylene-propylene;

    Filling liquid:silicone oil;

    Bolts:carbon steel zinc-plated, 1Cr18Ni9, 0Cr17Ni4CuNb, 42CrMo;

        Electrical housing:low-copper aluminum alloy;

        Coating:polyester epoxy resin;

        Pressure connection:For Range 3, 4, 5, the center hole distance of the two-flange connection thread 1/4-18NPT is 54mm;

         Connector 1/2-14NPT or M20X1.5 male thread;when equipped with a ball cone sealing connector, the center distance is 50.8, 54, 57.2mm;

        Weight:  3.5kg(excluding options)


                                                                                                       


Since its establishment, Xi'an Shenghongchuang Instrument Co., Ltd. has mainly been engaged in the R&D and production of pressure, level, temperature, differential pressure, digital display, weighing, and flow instruments. Main products include Swiss metallux ceramic piezoresistive sensors and self-developed ceramic capacitive pressure sensors. Diffused silicon, ceramic piezoresistive, strain gauge, metal capacitive, ceramic capacitive pressure transmitters, remote pressure transmitters, high-temperature pressure transmitters, flush diaphragm pressure transmitters, anti-corrosion pressure transmitters, integrated/split level transmitters, anti-corrosion level transmitters, lightning-protected level transmitters, high-temperature guided pressure level transmitters, straight rod level transmitters, flange level transmitters, wind pressure transmitters, metal capacitive differential pressure transmitters, diffused silicon differential pressure transmitters, pressure controllers, thermocouple/RTD temperature transmitters, and ultrasonic level gauges



Notes 1.All supplied products are provided with product certificates and instruction manuals, including product numbers, technical parameters, wiring methods, factory dates, etc. Please carefully check them to avoid misuse. 2.Wiring should be carried out strictly in accordance with our company's instructions for use. 3.This product is a precision instrument. Arbitrary disassembly, collision, dropping, or poking the sensor with sharp tools is prohibited. 4.The transmitter can work immediately after being powered on, but the output becomes stable after preheating for 30 minutes. 5.If any abnormality is found during use, the power supply should be turned off, use should be stopped, and inspection should be carried out, or contact our company's technical department directly. 6.During transportation and storage, restore the original packaging and store it in a cool, dry, and ventilated warehouse. 7.The quality warranty period is 12 months (damage caused by improper use such as pressure overload and incorrect wiring is not covered by the quality warranty). 8.Special products are customized products, do not participate in the 7-day no-reason return policy, and customized products have a delivery lead time, with no return or exchange.

u=2776578472,2395865860&fm=199&app=68&f=JPEG.jpg

017b2b6174636fa2a680162ff2d3055.jpg

1fb685ae1bbc54f4b2c3b2f1778e130.jpg

Submit

Related Recommendations