As a maintenance technician for flow control equipment, do you often encounter sudden failures in devices such as the 93718-353 electromagnetic flowmeter? This manual provides professional troubleshooting solutions for 7 common issues with insertion-type electromagnetic flowmeters with fluororubber lining, helping you quickly restore the normal operation of flowmeters in the HKTMF and JYLDE series and ensure the continuity of industrial production.
When HKTMF or JYLDE series electromagnetic flowmeters have no signal output, first check whether the power supply module is supplying power normally. Use a multimeter to measure the input terminal voltage and ensure it meets the 24VDC±10% range required for model TOP-101-32. If the power supply is normal, then check the following in sequence:
For split-type flowmeters such as XDL-DC, it is recommended to use an oscilloscope to detect the electrode signal waveform, which should normally show a linear characteristic of 2.5mV/(m/s).
When model 93636-135 electromagnetic flowmeters are used in the chemical industry, data fluctuations are often caused by the following reasons:
For the jump fluctuation issue of model 93718-353 in pulp flow measurement, field data from Xi'an Shenghongchuang Instrument Co., Ltd. show:
The empty pipe detection function of model 93454-543 electromagnetic flowmeters may trigger false alarms under the following conditions:
For TOP-101 series flowmeters, zero calibration should be performed according to the following steps:
Note: Model 142654-775 needs to be calibrated at an ambient temperature of 20±2℃, and every 10℃ temperature difference may cause a 0.5% error.
For XDL-DC electromagnetic flowmeters under strong acid working conditions, it is recommended to adopt the following measures:
When communication between model 93636-135 and the DCS system is interrupted, the following should be checked:
When the 93718-353 flowmeter installed near the pump room shows the following symptoms:
Solutions include installing rubber shock pads, replacing with armored cables, and tightening all mechanical connectors (torque must reach 2.5N·m).
Xi'an Shenghongchuang Instrument Co., Ltd. provides the following for electromagnetic flowmeters in the HKTMF and JYLDE series:
Contact our engineering team now to obtain a customized solution and ensure the continuous and stable operation of your flow measurement system.